Calibration PPM Meter and System OWS 15PPM Deckma OMD 2005 ASL Scorpio

Calibration PPM Meter and System OWS 15PPM Deckma OMD 2005 ASL Scorpio

Specification:

Equipment

PPM Meter

  • Brand DECKMA
  • Type OMD 2005
  • Range 0 – 30 ppm, Trend up to 50 ppm
  • Accuracy According IMO MEPC. 107(49)
  • Linearity Up to 30 ppm better than ± 2 %
  • Display Green Graphic Display
  • Power Supply 24 V – 240 V AC or DC Automatic Voltage Selection
  • Consumption < 15 VA
  • Alarm Points 1 + 2 Adjustable between 1 – 15 ppm (Works adjustment 15 ppm)
  • Alarm 1 Operating Delay (for annunciation purpose) Adjustable between 1 – 540 sec. (Works adjustment 2 sec)
  • Alarm 2 Operating Delay: (for control purposes) Adjustable between 1 – 10 sec. (Works adjustment 10 sec)
  • System Fault Alarm Red LED
  • Alarm Contact Rating Potential free 1 pole change over contacts, 3 A / 240 V
  • Alarm Indication Red LED’s
  • Output Signal 0 – 20 mA or 4 – 20 mA for 0-30 ppm reversible, ext. Load < 150 Ω
  • Sample Water Pressure 0,1 – 10 bar
  • Sample Flow Approx. 0,1 – 4 l/min depend. to pressure
  • Ambient Temperature + 1 to + 55° C
  • Sample Water Temperature + 1 to + 65° C
  • Roll Up to 45°
  • Size (over all) 360 mm W x 240 mm H x 100 mm D
  • Degree of Protection IP 65
  • Weight: 7,3 kg
  • Pipe Connections R ¼” Female

Main OWS

  • Brand Wehrmannsdann 11, D/27245KIRCHDORF Germany
  • Tested ACC to IMO Resolution MEPF 107 (49) TLEOYD 
  • Type Ocean Clean 0,5ED
  • Capacity 0,50 m3/H
  • Build 03.2007
  • Power 440V, 60Hz
  • Volume 0,16m3

Error Issue:

  1. Indications alarm sounds before the PPM meter showed a value of 15 PPM.
  2. Noisy pump motor.
  3. PPM meter detection results do not match.

Action Taken

  • PPM Meter calibration, result:

%
Miliampere
PPM
0
0
0
25
5
12.5
50
10
25
75
15
37.5
100
20
50

  • Alarm limit setting

Alarm 1
15PPM, Delay 2 Sec.
Alarm 2
15PPM, delay 10 Sec.
Offset setting
-3PPM
Ampere Output Range Setting
0 20 mA
  • Cleaning the glass tube detector.
  • Checking of water level.
  • Checking of Heater unit on Chamber.
  • Checking all automatic system, include solenoid valve for pneumatic control valve.

Solution Resulting:

  • PPM  Meter Testing has shown a corresponding value (standard 15 PPM).
  • Setting the alarm limits have been appropriate, results:
Not Alarm Condition in 10PPM
Not Alarm Condition in 10PPM
Alarm Condition in 16PPM
Alarm Condition in 16PPM
  • Motor noisy due to no water in the pump inlet. Testing the water sample.
  • All of the control system has been running as a standard, namely:
    • At a reading above 15 PPM, OWS output is returned to the beginning of the cleaning process. All settings are done by open and close the pneumatic valve.
    • Heating chamber worked well.
    • Water level detector works properly, the system ordered for running the pump motor when the water level in the low position. And stop the pump motor when the level in a high position.

Documentation

OWS Unit
OWS Unit
Chamber Heater and Water Level Sensor
Chamber Heater and Water Level Sensor
PPM Meter Unit
PPM Meter Unit
Main Panel OWS Unit
Main Panel OWS Unit